Safety

Emergency Stop

2.10 Emergency Stop
Note!
It is recommended to train the operator regularly in the use of the emergency stop function
so the operator feels confident in using it.
The System has six internal emergency stops; one connected to the overhead tube crane
and one on the servo button, one on each side of the table (at the head end) and two on the
wallstand.
Pressing one of the emergency stop buttons immediately cuts the power to all motorized
movements. The emergency stop is also connected to the generator. The emergency stop
prevents a new exposure and terminates an ongoing exposure.
A system message is displayed on the overhead tube crane display when the button is
activated.
To reset the emergency stop position, turn the emergency stop button clockwise. The button
is released and the system is ready for use again.
When the emergency stop is activated, lift the overhead tube crane up and push or pull
sideways to change the position of the tube/collimator manually.
There are additional external emergency stops as option.