To insert a new column stabilizer assembly:
1.
Firmly press the replacement column stabilizer assembly into the
thermal gasket. Ensure that the thermal gasket uniformly envelops the
sides of the column stabilizer and does not bunch beneath it.
Thermal gasket visible along
both sides of stabilizer
Threaded holes for
Phillips screws
Column stabilizer assembly (two styles):
For 150-mm column
Replacing the column stabilizer assembly
TP03425
For 50-mm or 100-mm column
Reusable fittings
Inlet tubing to injector
TP03123
September, 2019, 715003738IVD Rev. C
Column-stabilizer-
inlet tubing
75