Ferrule
3 mm
Collet
Older, reusable fittings do not have holes in their sides.
2.
Remove the column in-line filter unit from its packaging, and then push
the outlet end of the active preheater or column stabilizer tubing
assembly into the in-line filter inlet until it stops.
Installing the in-line filter:
Column
3.
Holding the reusable fitting in place, rotate the in-line filter unit onto
the fitting until it is snug.
Caution:
tighten the tubing fittings.
4.
Finger tighten the fitting until it is snug, and then tighten it up to an
additional 1/6-turn.
Reusable fitting
Flow
In-line filter unit
To avoid damaging the tubing, use only your fingers to
September, 2019, 715003738IVD Rev. C
Installing the optional in-line filter
Outlet tubing
From column stabilizer
tubing assembly
Inlet
Reusable fitting
TP02873
43