Column heater
Hide thumbs Also See for ACQUITY UPLC I-Class IVD:
Table of Contents

Advertisement

8.
Set the column heater aside.
9.
Ensure that the front left-hand foot of the column heater's rear
enclosure locks into the retainer bracket on the sample manager below
it.
If the retainer bracket is not installed on the sample
Alternative:
manager chassis, perform these tasks:
Warning:
solvent bottles from the solvent tray before removing the solvent
tray.
Caution:
remove the solvent bottles first. Then remove the solvent tray and,
finally, the modules, one at a time. Ask an assistant to help lift the
components off the system stack.
a.
Remove the solvent tray and any system modules stacked atop the
column heater's rear enclosure.
b.
Remove the Phillips screw that secures the column heater's rear
enclosure to the sample manager chassis.
c.
Remove the column heater's rear enclosure from the sample
manager's chassis.
d.
Secure the retainer bracket, supplied in the column heater start-up
kit, onto the chassis using the TORX screw provided in the kit.
To avoid injuries arising from solvent spills, remove the
To avoid spilling solvent when disassembling the system,
September, 2019, 715003738IVD Rev. C
Installing the column heater
Retainer bracket
27