Waters ACQUITY UPLC I-Class IVD Overview And Maintenance Manual page 47

Column heater
Hide thumbs Also See for ACQUITY UPLC I-Class IVD:
Table of Contents

Advertisement

6.
Attach the eCord fob to the receptacle on the right-hand side of the
device.
7.
Close the column compartment door, and then inspect the seal around
the front panel to ensure that it aligns correctly.
Proper alignment of the seal is critical for a stable thermal
Note:
environment.
To avoid inconsistent results due to temperature fluctuations,
Tip:
ensure that the column heater's or column manager's door latch is
completely closed.
Warning:
before you connect or disconnect the external communication cable
(between the sample manager and the column heater).
Caution:
• To avoid damage to the components, power-off the sample
manager before you connect or disconnect the external
communication cable (between the sample manager and the
column heater).
• To avoid damaging electrical components, do not disconnect an
electrical assembly from an instrument or device that remains
connected to a building's electrical supply. To completely
interrupt power, set the unit's power switch to Off. Then
disconnect the power cord from the unit's rear panel or from the
wall outlet. After discontinuing power, wait 10 seconds before
attempting to disconnect an assembly.
Heater compartment receptacle
To avoid electric shock, power-off the sample manager
September, 2019, 715003738IVD Rev. C
eCord fob
Installing the column
47