2 Preparing the column heater
Required materials
•
Clean, powder-free, chemical-resistant gloves
•
50/100-mm or 150-mm column stabilizer kit:
–
Collet and compression screw multi-tool (collet separator)
–
Column stabilizer assembly
–
PEEK tubing for connecting the column to the MS flow cell
–
Three Phillips screws (one is a spare)
–
Three thermal gaskets (two are spares)
Required tools
•
Needle-nose pliers
•
Phillips screwdriver
To install the optional column stabilizer assembly:
1.
Stop any currently running mobile phase before replacing the column
stabilizer assembly.
2.
Disconnect the inlet tubing from injector port 6 of the sample manager.
3.
Open the column heater door.
38
September, 2019, 715003738IVD Rev. C
Location of column-
stabilizer-inlet tubing