Maintenance Schedule - Waters ACQUITY UPLC I-Class IVD Overview And Maintenance Manual

Column heater
Hide thumbs Also See for ACQUITY UPLC I-Class IVD:
Table of Contents

Advertisement

4 Maintenance Procedures

Maintenance schedule

Perform the following routine maintenance on the column heater to ensure
reliable operation and accurate results. When using the system throughout
the day (and on nights and weekends), or when using aggressive solvents,
such as buffers, perform the maintenance tasks more frequently.
For information about scheduling routine preventative maintenance based on
system usage, see
Recommended routine maintenance schedule:
Maintenance
procedure
Replace the
leak sensor on
the column
heater
Replace the
column
58
September, 2019, 715003738IVD Rev. C
page
63.
Frequency
As needed. For example, when the leak
sensor is damaged or an error message
states that the leak sensor is not detected.
As needed. The typical failure modes are as
follows:
• High back-pressure. The column causes a
high-pressure shutdown.
• Method requirements are not attained.
This can be exhibited by poor peak shape,
resolution between peaks lost, tailing,
and so on. Perform a plate count to
establish column efficiency.
For
information
See
page
63.
See
page
69.