B Specifications
Performance specifications: (Continued)
Item
Column compartment temperature
stability
Solvent conditioning
Wetted materials of construction
The following table lists the wetted materials of construction for the
ACQUITY UPLC I-Class IVD column heater:
Wetted materials of construction:
Description
Wetted materials
98
September, 2019, 715003738IVD Rev. C
Specification
Tested to ±0.3 C
Test conditions:
• Door closed
• No column installed
• No flow
• Measurement taken with traceable,
external temperature measurement
device
• Measurement taken after one hour
of thermal equilibration at setpoint
• Measurement taken at column
compartment sensor location
Tested at 35 °C, 55 °C, and 85 °C
• Active preheating as standard
• Passive preheating (also
recommended in CH-A only for
legacy method support)
Specification
316L stainless steel