Process for installing the MS-bracket
Task
Remove the column heater from the device chassis
Install the MS-bracket on the column heater chassis
Attach the column heater housing onto the MS-bracket
Install the column heater and connecting accessories, such as
the MS drip tray, leak sensor, and drain tubing
Install the column and connect inlet and outlet tubing

Required materials

Column heater MS-bracket kit
MS-bracket
Column heater housing
MS drip tray
Hinge assembly
Cable clamp
®
TORX
screws
®
Phillips
Cable clamp nut
Drain tubing
Clean, powder-free, chemical-resistant gloves

Required tools

1/4-inch open-end wrench
4-mm open-end wrench
3-mm hex wrench
Phillips screwdriver
Small, flat screwdriver
T20 TORX driver
screws
September, 2019, 715003738IVD Rev. C
Installing the column heater
See
page 24
page 25
page 29
page 30
page 34