Process for installing the MS-bracket
Task
Remove the column heater from the device chassis
Install the MS-bracket on the column heater chassis
Attach the column heater housing onto the MS-bracket
Install the column heater and connecting accessories, such as
the MS drip tray, leak sensor, and drain tubing
Install the column and connect inlet and outlet tubing
Required materials
•
Column heater MS-bracket kit
–
MS-bracket
–
Column heater housing
–
MS drip tray
–
Hinge assembly
–
Cable clamp
®
–
TORX
screws
®
–
Phillips
–
Cable clamp nut
•
Drain tubing
•
Clean, powder-free, chemical-resistant gloves
Required tools
•
1/4-inch open-end wrench
•
4-mm open-end wrench
•
3-mm hex wrench
•
Phillips screwdriver
•
Small, flat screwdriver
•
T20 TORX driver
screws
September, 2019, 715003738IVD Rev. C
Installing the column heater
See
page 24
page 25
page 29
page 30
page 34
23