c) Re-measurement of the same patient

3.6.7 Checking measurement data

a) Intraocular pressure correction
3-52
Step on the foot switch to capture measurement data. A beep sounds when the
measurement data is captured.
Repeat measurements in the same manner. When 10 sets of measurement data
are taken, a beep sounds to indicate completion of measurement.
To take measurement data in the next memory slot before 10 sets of
measurement data have been taken, touch the next memory number to enter
standby mode and perform measurement.
When the pachymetry probe is removed from the cornea, the next memory slot
enters standby mode.
The deleted data cannot be restored. Carefully check the data before
deleting it.
Retain the patient information and delete only the measurement data for each eye.
Delete all data referring to "3.1.5 Deleting all measurement data (measurement
preparation for other patient)" when the patient is changed.
Touch the "Retake" button (1) for approximately 1 second until a beep sounds to
delete the data. The patient information is retained and the measurement data of
the eye other than the inspected eye being displayed.
This instrument automatically starts calculation and displays the result when all
items required for calculating the intraocular pressure correction are set.
(1)
(1)
(Fig. 1)
(3)
(4)
(5)
(6)
(2)
(7)