Tomey UD-800 Instruction Manual page 124

Ultrasonic a/b scanner and pachymeter
Table of Contents

Advertisement

Setting
Example
1) Check the computer IP address.
Check and record the IP address and subnet mask of the computer with DATA
Transfer installed. Refer to the DATA Transfer operation manual for details.
2) Settings on the Instrument
The settings shown in the table below describe an example when the IP address
of the computer with DATA Transfer installed is "192.168.2.128" and the subnet
mask is "255.255.255.0." The IP setting method in this case is "Manual."
(Local)
IP address
Subnet
mask
Default
gateway
Host IP
address
*1: An arbitrary number from 1 to 255 excluding numbers used by the computer
(128 in this example)
"CˆÓ‚Ì "Žš
3-92
■ When only this instrument and the Tomey Link Server or a
computer running Data Transfer are connected to the network,
the following settings can be used.
Computer settings
192
168
2
Check these on the DATA
Transfer screen.
255
255
255
Check these on the DATA
Transfer screen.
-------
-------
Instrument settings
128
192
168
Same value as
computer setting
0
255
255
Same value as computer
setting
0
0
All "0"
192
168
IP address of computer
2
129
(*1)
255
0
0
0
2
128