5.5 Replacing consumables

5.5.1 Fuses

Disconnect the power cord from the outlet when replacing fuses.
Otherwise you may get electric shock, resulting in death or serious
injuries.
Use fuses specifically designed for this instrument.
When the instrument does not work correctly after fuses have been
replaced, there may be other causes of the problem. Turn off the
instrument immediately and contact your local distributor.
1)
Turn power off and disconnect the power plug from the outlet.
2)
Disconnect the power plug from the power terminal.
3)
Insert a flat screwdriver into the groove of the fuse holder under the power
terminal and remove the holder case.
4)
Replace the blown fuse with a new one.
5)
Install the fuse holder in the reverse order of removal.

5.5.2 Printer paper

Always use genuine TOMEY paper for the printer. Using other types of
paper may cause printer failure.
Do not pull paper forcibly. Trying to pull the paper out may cause
printer failure. Touch the "PRINT" button without any data stored to
feed blank paper.
Replace the paper roll according to the procedure below when red lines appear
on the edges of the printer paper.
1)
Press the button (1) to open the printer cover.
2)
Remove the old printer paper roll and install a new roll. If the paper roll is
(1)
installed in the wrong direction, nothing will be printed.
3)
Close the printer cover with the end of the paper protruding 2 mm from the
outlet. Press the cover firmly until you hear a click.
5-7