Declaration Of Conformity To Emc - Tomey UD-800 Instruction Manual

Ultrasonic a/b scanner and pachymeter
Table of Contents

Advertisement

8.6 Declaration of Conformity to EMC

When using specified accessories or cables excluding internal parts sold as
repair parts of TOMEY Corporation, electromagnetic emission of the UD-800 may
increase or electromagnetic immunity may decrease.
When using handheld or portable communication equipment, locate it more than
30cm(12inches) away from the UD-800. It may degrade the performance of the
UD-800 if the equipment is located closer.
Do not use the UD-800 adjacent to or stacked on other equipment. If it needs to be
adjacent to or placed on other equipment, it must be confirmed that UD-800
operates correctly in such location.
The UD-800 falls under Group 1, Class B according to EN 55011(CISPR 11). This
means that the UD-800 does not generate RF energy intentionally in the form of
electromagnetic irradiation, inductive coupling and/or capacitive coupling for
processing of materials or inspection/analysis, and that it is suitable for use in
facilities directly connected to a low-voltage power network that supplies power to
general household facilities and buildings used for home use.
The UD-800 can be used in a specialized health care facility such as hospitals and
clinics (except anywhere near the source of a strong EMI) and a general
commercial facility such as eye glasses stores.
The UD-800 requires special attention for EMC and needs to be installed, serviced
and used based on the following information.
Do not use cables other than those provided or specified by TOMEY Corporation.
The person who connects additional equipment to the signal I/O section as part of
the medical system, shall take responsibility for ensuring the system conforms to
IEC/EN 60601-1-2 requirements.
8-9