Troubleshooting; Common Items - Tomey UD-800 Instruction Manual

Ultrasonic a/b scanner and pachymeter
Table of Contents

Advertisement

6. TROUBLESHOOTING

6.1 Common items

Check the following first when you encounter any problems.
If the problem is not solved even after checking the applicable item listed
below, contact your local distributor to request inspection and/or repair.
Do not remove the cover of the instrument. Otherwise you may get
electric shock.
Do not take any actions other than those specified below.
 The instrument does not start when the power switch is turned on.
 Nothing appears on the monitor screen.
Problem with the power plug
Check that the power plug is firmly connected to the
outlet. Check that there are no flaws in the power cord,
such as cracks or tears.
Problem with the power outlet
Check that power is supplied to the outlet to which the
power cord is connected.
Blown fuse
Check that fuses are not blown. If blown, replace the
fuse (refer to "5.5.1 Fuses"). When the new fuse is
blown again, the instrument may be faulty. Contact
your local distributor to request inspection and/or
repair.
The auto power off function, which automatically turns
off the screen when the instrument is not operated for
the specified time, has been activated.
Touch the monitor screen.
6-1