Tomey UD-800 Instruction Manual page 121

Ultrasonic a/b scanner and pachymeter
Table of Contents

Advertisement

(8) SCU AE Name
Title of DICOM Application Entity.
Conforms to the DICOM standard, and is used for communicating
connection between AEs for bilateral certification.
When registered on the server (SCP), unauthorized access is prevented,
and registered users can identify this application on the network.
(9) SCP AE Name
Set the title of the DICOM Application Entity compatible with the DICOM
server (SCP).
DICOM communication is generally conducted by combinations of IP
address, port number and AE title. Be sure to set this SCP AE name.
Check the DICOM Conformance Statement before setting the AE title of the
communication destination.
(10) SCP IP Address
Set the IP address of the DICOM compatible server (SCP).
(11) Port Number
Set the port number used for DICOM communication.
(12) "Connection Test" button
Sends a "C-ECHO-RQ" message to the DICOM compatible server (SCP).
During DICOM communication, C-ECHO service is conducted to check
connection between the server and terminal ends.
3-89

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents