i.
Important safety information
Do not install this instrument in a location where explosives or flammable
substances are used or stored. Otherwise, fire or explosion may occur.
Do not remove the cover of the instrument. Otherwise electric shock may
result.
Do not disassemble or modify the instrument. Otherwise electric shock may
result.
Disconnect the power cord from all the connected devices before installing
and/or servicing the instrument. Otherwise, electric shock may result.
Sterilize probes and the eye cup before use.
Do not use probes and the eye cup with its contact section damaged.
Measurement error may occur, as well as the cornea or eyelid may be injured.
Do not commence service or maintenance work while the instrument is being
used for a patient.
Do not place water or chemicals on the instrument. Any water or chemicals
entering the instrument may cause electric shock or failure.
Do not allow water or chemicals to splash on probes, except for the tip cap.
Any water or chemicals entering the instrument may cause electric shock or
failure.
Use only the specified data terminal for connection of the instrument. Using
another type of terminal may result in failure of the instrument.
This instrument is a diagnostic/measuring device specially designed for
ophthalmology. Never use the instrument for other purposes.
The terminal for connecting the instrument to external devices is not isolated
from the internal circuit. Inappropriate wiring may damage the internal circuit.
Never touch any of these terminals and the patient at the same time. Be sure to
contact Tomey or your local distributor before using the external output
terminal.
When operating this instrument connected to other devices, use only devices
that satisfy IEC60601-1 or equivalent safety requirements, or that conform to
IEC60950 and whose power source is isolated with an isolation transformer, in
order to satisfy the relevant safety requirements for medical equipment.