4.1.3 Holladay 1

1. Implanted IOL power (D)
1000
na
P
L
2
C
2
SF
2. Predicted refractive power after surgery (D)
REF
iol
na
V
X
Where:
X
: na・r – L2 (nc – 1)
Y
: na・r - (nc - 1)(C2 + SF)
Q
: L2 – C2 -SF
na
: Aqueous humor and vitreous refractive index
= 1.336
nc
: Corneal reflective index
= 4/3
L
: Axial length (mm)
K
: Average corneal refractive power (D)
= (K1 + K2)/2
r
: Radius of corneal curvature (mm)
= {(KI - 1) × 1000}/K
KI
: Cornea equivalent refractive index
SF
: Distance from the iris surface to the optical center of the
implanted IOL (mm)
REF : Target Ref. after surgery (D)
V
: Vertex distance (mm)
= 12
P
: Implanted IOL power (D)
L2
: Corrected axial length (mm)
= L + 0.2
C2
: Anatomic anterior chamber depth; distance from the
corneal vertex to the iris surface (mm)
= 0.56 + Rag –
r < 7
→ Rag = 7
r ≧ 7
→ Rag = r
X
. 0
001
REF
V
X
Y
. 0
001
REF
V
Y
1000
na
X
P
Q
Y
L
2
r
. 0
001
P
Q
V
Y
Rag 
2
2
AG
/
L
2
r
r
C
2
SF
r
C
2
SF
4
4-5