4.1.6 SRK SHOWA formula

1. Implanted IOL power (D)
L < 22.0
:P = A – 2.5L – 0.9K + 1.4 – 1.45REF
22.0≦ L < 24.5
: P = A – 2.5L – 0.9K – 1.67REF
24.5≦ L < 27.0
: P = A – 2.5L – 0.9K + 0.71 – 1.25REF
27.0≦ L
: P = A – 2.5L – 0.9K – 1.69REF – 1.69
2. Predicted refractive power after surgery (D)
L < 22.0
: R = (A - 2.5L – 0.9K + 1.4 – P)/1.45
22.0≦ L < 24.5
: R = (A – 2.5L – 0.9K – P)/1.67
24.5≦ L < 27.0
: R = (A – 2.5L – 0.9K + 0.71 – P)/1.25
27.0≦ L
: R = (A – 2.5L – 0.9K – 1.69 – P)/1.69
L
: Axial length
K
: Average corneal refractive power (D)
= (K1 + K2)/2
r
: Radius of corneal curvature (mm)
= {(KI – 1) × 1000}/K
KI
: Cornea equivalent refractive index
REF : Target Ref. after surgery (D)
A
: A-Constant
P
: Implanted IOL power (D)
3. Personal A-Constant
L < 22.0
: A = P + 2.5L + 0.9K - 1.4 + 1.45AREF
22.0≦ L < 24.5
: A = P + 2.5L + 0.9K + 1.67AREF
24.5≦ L < 27.0
: A = P + 2.5L + 0.9K - 0.71 + 1.25AREF
27.0≦ L
: A = P + 2.5L + 0.9K + 1.69AREF + 1.69
AREF : Eye refractive power after surgery (D)
4-11