Biometry / IOL power calculations
●Measurement range
●Accuracy
●Factory setting of the converted velocity
8-2
Axial length
ACD
Lens thickness
Measurement Accuracy
Measurement Resolution
Average velocity of the axial length
Phakic Eye
Dense Cataract
Aphakic
Lens velocity
Phakic Eye
Dense Cataract
ACD velocity
Phakic Eye
Dense Cataract
IOL Eye
Velocity of vitreous
Phakic Eye
(In case of sectional velocity)
IOL Eye
IOL velocity
Acrylic
Silicone
PMMA
User
IOL thickness
Acrylic
Silicone
PMMA
User
: 15.00 - 45.00mm
: 1.80 - 7.00mm
: 2.00 - 6.00mm
: ±0.1mm
: 0.01 mm
: 1,550m/s
: 1,548m/s
: 1,532 m/s
: 1,641m/s
: 1,629m/s
: 1,532 m/s
: 1,532 m/s
: 1,532 m/s
: 1,532 m/s
: 1,532 m/s
: 2,200m/s
: 1,049m/s
: 2,718m/s
: 2,200m/s
: 0.80mm
: 1.00mm
: 0.80mm
: 0.80mm