3.2.5 Obtaining ultrasonic image

3-14
Do not allow water or chemicals to splash on probes, except for the tip
cap. Any water or chemicals entering the instrument may cause electric
shock or failure.
Be sure that the screw is not loose when you use the arm. Otherwise
the probe attached to the arm may drop to be damaged or may drop to
damage an eye ball or eye lid of the patient. Operating the arm directly
with your hand may injure your finger pinched by the arm.
Be sure to touch the "New" button to delete all the measurement data
of the previous patient before measuring another patient.If new
measurement is started without deleting the previous data, the
measurement data of the previous patient may be included.
Releasing a freeze by touching the FREEZE button or stepping on the
foot switch deletes all images previously obtained.Be very careful when
performing initialization.
(1)
1) When the instrument is in FREEZE mode, touch the "FREEZE cancel" button
(1) or step on the foot switch to release FREEZE mode.
2) Apply an appropriate amount of the ultrasound diagnosis gel to the contact
section of the B mode probe.
3) Apply the B mode probe directly to the lid of the examined eye.