5
Maintenance and Care

Cleaning the Touchscreen

Cleaning the Touchscreen
1. Pull out the power supply plug.
2. Clean the touchscreen using a dry microfiber cloth.
3. If necessary moisten the microfiber cloth and wipe the touchscreen again.

Cleaning the Condenser Filter

The Sorvall LYNX 4000 / 6000 centrifuge has two filter mats to prevent dust from entering the
centrifuge. In order to clean them proceed as follows:
1. Loosen the four screws of both venting grids on the right side of the centrifuge.
2. Remove the venting grids and the filter mats (75000014).
3. Use a vacuum cleaner on both sides of both filter mats.
4. Inserting the filter mats again.
5. Reinstall the venting grids.
Figure 17. Venting Grids
38
Sorvall LYNX 4000 / 6000
CAUTION The fins of the condenser are very sharp. Make sure to wear protective gloves
when removing the filter mats.
Screws
Filter Mats