Technical Data - Thermo Scientific Sorvall LYNX 4000 Instruction Manual

Superspeed centrifuge
Hide thumbs Also See for Sorvall LYNX 4000:
Table of Contents

Advertisement

Technical Data

The technical data of the Sorvall LYNX 4000 / 6000 centrifuge is listed in the following table.
Table 1. Technical Data
Feature
Environmental Conditions
Environmental Conditions
during Storage and Shipping
Permissible Ambient Temperature
Overvoltage Category
Pollution Degree
Heat Dissipation /h
IP (statement of the degree of protection
according to IEC 60529)
Run Time
Maximum Speed n
Minimum Speed n
Maximum RCF-Value at n
Maximum kinetic Energy
Noise Level for swinging Bucket at
maximum Speed
Noise Level for Fixed Angle Rotors at
maximum Speed
Temperature Setting Range
Dimensions
Height with Door closed (incl. GUI)
Height with Door open
Width
Depth
Weight without Rotor
* Rotor A27-8x50, Typical Run 4 °C at 24,000 rpm 30 minutes, 60 Hz +5 %, 30 min stand by 4° C (lid closed)
** Measures at 1 m distance and at 1.6 m height
Thermo Scientific
*
max
min
max
Sorvall LYNX 4000
-Use in interior Spaces
-Altitudes of up to 2,000 m above Sea Level
-Max. relative humidity 85 % up to 30 °C
Temperature:
-10 °C to +55 °C
Humidity: 15 % to 85 %
+5 °C to +35 °C
II
2
2.0 kWh / 6,830 Btu / 7,200 kJ
20
99 hours:59min, hold
24000 rpm (depending on
the rotor)
500 rpm
68905 x g
< 203 kJ
**
< 61 dB (A)
**
< 59 dB (A)
-10 °C to +40 °C
1048 mm
1531 mm
700 mm
805 mm
285 kg
1
Introduction and Description
Technical Data
Sorvall LYNX 6000
Temperature:
-10 °C to +55 °C
Humidity: 15 % to 85 %
+5 °C to +35 °C
II
2
1.2 kWh / 4,100 Btu / 4,320 kJ
20
99 hours:59min, hold
29000 rpm (depending on
the rotor)
500 rpm
100605 x g
< 203 kJ
**
< 61 dB (A)
**
< 59 dB (A)
-20 °C to +40 °C
1048 mm
1531 mm
700 mm
805 mm
295 kg
Sorvall LYNX 4000 / 6000
3

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sorvall lynx 6000

Table of Contents