The following table contains an overview on the electrical contact data of the
Sorvall LYNX 4000 / 6000 centrifuge. This data is to be taken into consideration when selecting the
mains connection socket.
Table 4. Electrical Connection Data
Mains Voltage
in V
200
208
220
230
240
380
400
415
200
208
220
230
240
380
400
415
1
For the 200-240 V units the inrush current is 120 A for up to 1 second during the refrigeration start period. For the 380 / 400 V units it is 60 A. Circuit breakers,
whether thermal or magnetic actuated, must have a delay type applicable for the start of motors.
2
Use a 25 A or 32 A Trip Char. C circuit breaker (D or K are also applicable).
3
For 380 / 400 V, 3-phase (unbalanced load, no neutral used) use a 16 A Trip Char. C circuit breaker (D or K are also applicable).
4
For North America: use for example GES-9888 30 A.
6
Sorvall LYNX 4000 / 6000
Frequency
Rated Current
in Hz
in A
50
24
50
23
50
21
50
20
50
19
50
13.5
50
12.5
50
11.5
60
24
60
23
60
21
60
20
60
19
60
13.5
60
12.5
60
11.5
Power Consump-
Building Fuse in
1
tion in W
A
4600
30
4600
30
4600
32
4600
32
4600
32
4600
16
4600
16
4600
16
30
4800
4800
30
4800
32
4800
32
4800
32
4800
16
4800
16
16
4800
Equipment
Fuse in A
30
30
2
30
2
30
2
30
3
16
3
16
16
3
30
30
2
30
2
30
2
30
3
16
3
16
16
Thermo Scientific