Application instructions

3
Application instructions
Piezoelectric force washers of the CLP type series are suitable for measuring
compressive forces. Because they provide highly accurate dynamic and quasi-
static force measurements, they must be handled very carefully. Particular care
must be taken when transporting and installing the devices. Dropping or knock­
ing the transducer may cause permanent damage.
The transducers are extremely stiff and have a high natural frequency.
Limits for acceptable mechanical, thermal and electrical stress are listed in the
Specifications section on page 19. It is essential that these are taken into ac­
count when planning the measuring set-up, during installation and, ultimately,
during operation.
Notice
Like all piezoelectric force sensors, the piezoelectric force washers of the CLP
series are only exposed to great force (greater than 10% of the sensor capac­
ity) if they are connected to a charge amplifier, or short-circuited. If the sensors
are loaded without this measure being taken, voltage flashover can result,
which could damage the sensor.
CLP
9
A03616_04_YI0_02 HBM: public