TECHNICAL DATA

DESCRIPTION
Manufacturer
Model
Classification EN 60601-1
Classification 93 / 42 EWG
Mode
with irrigation:
without irrigation:
Supply voltage
Power consumption
Fuses
Water supply
Compressed air supply
Ultrasonic specifications
Maximum output:
Frequency range:
Weight
Dimensions in mm (HxWxD)
Operating conditions
Storage and
transport conditions
Date of manufacture
Life time
68
FB-305_CN_rev_C_AFS2.indd 68
EMS SA, CH-1260 Nyon, Switzerland
AIR-FLOW
S2
®
- Class I
- Applied part, Type B
- IP 20 unit
- IP X1, footpedal
Class IIa
Continuous operation
Duty cycle 10 % during 10 min. maximum
100 -110 VAC / 50 / 60 Hz
120 - 130 VAC/ 50 / 60 Hz
220 VAC/ 50 / 60 Hz
230 - 240 VAC / 50 / 60 Hz
max. 80 VA
Ø 5 x 20 mm, 250 VAC, time lag, low
breaking capacity
for 100 - 130 VAC power supply: T1AL
for 220 - 240 VAC power supply: T0.5AL
1 to 5 bar (100 to 500 kPa)
4.5 to 7 bar (450 to 700 kPa)
8 W
25 - 32 kHz
Approx. 4.0 kg
120 x 250 x 250
+10° C to +40° C
30% to 75% relative humidity
Altitude max. 3000 m
-10°C to +40°C
10% to 95% relative humidity
500 hPa to 1060 hPa air pressure
See the packaging
7 Years