[:Sense]:Frequency:stop; [:Sense]:Power[:Rf]:Range - Rigol SA5000 Series Programming Manual

Vector signal analysis
Table of Contents

Advertisement

RIGOL

[:SENSe]:FREQuency:STOP

Syntax
[:SENSe]:FREQuency:STOP <freq>
[:SENSe]:FREQuency:STOP?
Description
Sets the stop frequency.
Queries the stop frequency.
Parameter
Name
<freq>
Consecutive Real Number
Return Format
The query returns the stop frequency in scientific notation. The unit is Hz.
Example
The following command sets the stop frequency to 10 MHz.
:SENSe:FREQuency:STOP 10000000
The following query returns 1.000000000e+07.
:SENSe:FREQuency:STOP?

[:SENSe]:POWer[:RF]:RANGe

Syntax
[:SENSe]:POWer[:RF]:RANGe <real>
[:SENSe]:POWer[:RF]:RANGe?
Description
Sets the amplitude of the largest sinusoidal input signal without being clipped by the IF ADC.
Queries the amplitude of the largest sinusoidal input signal without being clipped by the IF ADC.
Parameter
Name
<real>
Return Format
The query returns the signal amplitude in scientific notation. The unit is dBm.
Example
The following command sets the amplitude of the largest sinusoidal input signal without being clipped by
the IF ADC to 20 dBm.
:SENSe:POWer:RF:RANGe 20
The following query returns 2.000000000e+01.
:SENSe:POWer:RF:RANGe?
2-60
Type
3.125 kHz to Fmax
Type
Integer
-15 dBm to 25 dBm
Range
center frequency + span/2
Range
Chapter 2 Command System
Default
Default
20 dBm