[:Sense]:Ddemod:filter:measurement:user<N>[:Data]; [:Sense]:Ddemod:filter:reference - Rigol SA5000 Series Programming Manual

Vector signal analysis
Table of Contents

Advertisement

RIGOL
[:SENSe]:DDEMod:FILTer:MEASurement:USER<n>[:DATA]
Syntax
[:SENSe]:DDEMod:FILTer:MEASurement:USER<n>[:DATA] <ampl>{,<ampl>}
[:SENSe]:DDEMod:FILTer:MEASurement:USER<n>[:DATA]?
Description
Sets the data of the specified measurement user-defined filter.
Queries the data of the specified measurement user-defined filter.
Parameter
Name
<n>
Discrete
<ampl>
Consecutive Real Number
Return Format
The query returns the user-defined filte data in scientific notation.
Example
The following command sets the data of the measurement user-defined filter 1.
:SENSe:DDEMod:FILTer:MEASurement:USER1:DATA
-0.000321525,0.00294817,0.002500879,-0.001910775,-0.005052538
The following query returns
-3.215250000e-04,2.948170000e-03,2.500879000e-03,-1.910775000-03,-5.052538000e-03。
:SENSe:DDEMod:FILTer:MEASurement:USER1:DATA?

[:SENSe]:DDEMod:FILTer:REFerence

Syntax
[:SENSe]:DDEMod:FILTer:REFerence RCOSine|RRCosine|GAUSsian|RECTangle|HSINe|USER<n>
[:SENSe]:DDEMod:FILTer:REFerence?
Description
Sets the reference filter type.
Queries the reference filter type.
Parameter
Name
Type
--
Keyword
<n>
Discrete
Remarks
RCOSine: indicates the raised cosine filter.
RRCosine: indicates the root raised cosine.
GAUSsian: indicates the Gaussian filter.
RECTangle: indicates the rectangular filter.
HSINe: indicates the half-sine filter.
USER<n>: indicates the user-defined filter n (1 to 6).
Return Format
The query returns RCOS, RRC, GAUS, RECT, HSIN, USER1, USER2, USER3, USER4, USER5, or USER6.
Example
The following command sets the reference filter to GAUSsian.
2-48
Type
RCOSine|RRCosine|GAUSsian|RECTangle|HSINe|USER
Range
1|2|3|4|5|6
——
Range
1|2|3|4|5|6
Chapter 2 Command System
Default
——
——
Default
RCOSine
——
VSA Programming Guide

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rsa5065Rsa5032Rsa5065-tgRsa5032-tg

Table of Contents