Checking Proper Connection of SMP Connectors
AWG7000 Series Service Manual
The cable at J690 from the CH2 OUT10G board to the Channel 2 Analog
connector.
The cable at J790 from the CH2 OUT10G board to the Channel 2 Analog
connector.
The cable at J120 from the CH2 OUT10G board to the Channel 2 Mkr 1
connector.
The cable at J130 from the CH2 OUT10G board to the Channel 2 Mkr 1
connector.
The cable at J160 from the CH2 OUT10G board to the Channel 2 Mkr 1
connector.
The cable at J190 from the CH2 OUT10G board to the Channel 2 Mkr 1
connector.
7. Remove the four screws securing the CH2 OUT10G board to the chassis.
8. Lift the board away from the chassis.
Installation. To install, reverse this procedure.
After you have replaced the AWG10G board or OUT10G board(s), be sure to verify
that the center contacts of the SMP cable connectors do not short the ground
terminals at J100, J110, J150, and J170 on the OUT10G board(s).
You need a digital multimeter to check the proper connection.
Perform the following steps:
1. Set the digital multimeter to the
2. Attach the red test lead to the transmission line near J100 on the OUT10G
board. See Figure 4-17.
3. Attach the black test lead to the connector housing at J100 on the OUT10G
board. See Figure 4-17.
4. Check that the digital multimeter reading falls within the range of 30
70
.
5. Repeat steps 2 through 4 for J110, J150, and J170.
6. If your instrument has two OUT10G boards, repeat steps 2 through 5 for the
other board.
Removal and Installation Procedures
mode.
to
4-37