Procedures for Internal Modules (Lower)
Front Connector Board
AWG7000 Series Service Manual
You need a screwdriver with a T-15 Torx tip (items 1 and 2).
Removal. To remove the Front Connector board, refer to Figure 4-13 on page 4-28
and follow these steps:
1. Remove the handle unit, snaps, cosmetic covers, front-trim unit, and EMI
covers.
2. Orient the instrument so that the top is on the work surface.
3. Disconnect the cables from J100 and J200 on the Front Connector board.
4. Remove the screw securing the Front Connector board to the chassis.
5. Grasp the board and pull it gently away from the chassis.
Installation. To install, reverse this procedure.
Removal and Installation Procedures
4-27