Tektronix AWG7000 Series Service Manual page 76

Arbitrary waveform generators
Hide thumbs Also See for AWG7000 Series:
Table of Contents

Advertisement

Removal and Installation Procedures
Power Supply
RFI Filter
4-24
You need a screwdriver with a T-15 Torx tip (items 1 and 2).
Removal. To remove the power supply, refer to Figure 4-11 and follow these steps:
1. Remove the handle unit, snaps, cosmetic covers, front-trim unit, EMI covers,
drive-bay module, and Multi-Input/Output board.
2. Orient the instrument so that the bottom is on the work surface.
3. Remove the six screws securing the power supply bracket to the chassis.
4. Disconnect the cable from J611 on the Power board.
5. Remove the three screws securing the bracket to the power supply.
6. Carefully slide the module to loosen it from the Power board.
7. Lift the power supply up out of the chassis.
8. Lift the bracket up out of the chassis.
Installation. To install, reverse this procedure.
You need a screwdriver with a T-15 Torx tip (items 1 and 2).
Removal. To remove the power supply, refer to Figure 4-11 and follow these steps:
1. Remove the handle unit, snaps, cosmetic covers, and EMI covers.
2. Orient the instrument so that the bottom is on the work surface.
3. Disconnect the cable from J600 on the Power board.
4. Remove the two screws securing the ground leads to the chassis.
5. Remove the two screws securing the RFI filter to the chassis.
6. Pull the RFI filter away from the chassis.
Installation. To install, reverse this procedure.
AWG7000 Series Service Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

Awg7101Awg7102Awg7051Awg7052

Table of Contents