Removal and Installation Procedures
4-32
Removal. To remove the relay unit, refer to Figure 4-13 on page 4-28 and follow
these steps:
1. Remove the handle unit, snaps, cosmetic covers, front-trim unit, EMI covers.
2. Orient the instrument so that the top is on the work surface.
3. Use a 5/16 inch torque wrench to remove the 14 SMA cables from the relay
units.
4. Disconnect the four cables from the relay units.
5. Remove the 10 screws securing the relay unit to the chassis.
6. Lift the relay unit away from the chassis.
Installation. To install, reverse this procedure. See Figure 4-16 for cable
connection.
To J1000 on the CH2 OUT10G board
To J1002 on the CH2 OUT10G board
Front
To the Interleave Analog connector
To the Interleave Analog connector
To J1002 on the CH1 OUT10G board
To J1000 on the CH1 OUT10G board
Figure 4-16: Relay unit cable connection for option 06
To J1012 on the CH2 OUT10G board
To the Channel2 Analog connector
To J1010 on the CH2 OUT10G board
To the Channel2 Analog connector
To J101B on the AWG10G board
To J100B on the AWG10G board
To J101A on the AWG10G board
To J100A on the AWG10G board
To the Channel1 Analog connector
To J1012 on the CH1 OUT10G board
To the Channel1 Analog connector
To J1010 on the CH1 OUT10G board
AWG7000 Series Service Manual