Tektronix AWG7000 Series Service Manual page 106

Arbitrary waveform generators
Hide thumbs Also See for AWG7000 Series:
Table of Contents

Advertisement

Troubleshooting
Table 4-8: Hardware error (cont.)
Error code
Error log message
300000012
The Ipc busy timeout period has expired.
300000013
MemoryController DLL doesn't lock.
300000014
MemoryController Dynamic Phase Shift - IPC error.
300000015
MemoryController Dynamic Phase Shift - Waveform RAM error.
300000016
MemoryController Dynamic Phase Shift - System Clock error.
300000017
The Diag busy timeout period has expired.
300000018
MemoryController Dynamic Phase Shift - Mux error.
300000019
AWG10G board Calibration checksum data has an error.
300000020
AWG10G board Calibration version unmatched.
300000100
The access of MIO board EEPROM, it failed.
300000101
The access of OUTPUT board EEPROM, it failed.
300000102
The relay count is invalid. Board-1
300000103
The relay count is invalid. Board-2
300000104
The access of AWG board EEPROM, it failed.
300000200
The download (INIT) timeout period has expired.
300000201
The download (INIT) timeout period has expired.
300000202
The download (DONE) timeout period has expired.
300000203
The download timeout period has expired.
300000204
The system cannot find the file specified.(PLD140b.Bit)
300000205
An attempt was made to load a program with an incorrect
format.(PLD131.Bit)
300000206
The download(AwgController-INIT) timeout period has expired.
300000207
The download(AwgController-INIT) timeout period has expired.
300000208
The download(DONE) timeout period has expired.
300000209
The download timeout period has expired.
300000210
The system cannot find the file specified.(PLD131.Bit)
300000211
An attempt was made to load a program with an incorrect
format.(PLD131.Bit)
300000212
The download (AwgController-INIT) timeout period has expired.
300000213
The download (AwgController-INIT) timeout period has expired.
300000214
The download (AwgController-DONE) timeout period has expired.
300000215
The download (AwgController) timeout period has expired.
300000216
The system cannot find the file specified.(PLD130b.Bit)
4-54
Possible defect
Note
on board
AWG10G
AWG10G: U1000
AWG10G
AWG10G: U1000, U400xx
AWG10G
AWG10G: U1000, U400xx
AWG10G
AWG10G: U1000, U400xx
AWG10G
AWG10G: U1000, U400xx
AWG10G
AWG10G: U1000, U400xx
AWG10G
AWG10G: U1000, U400xx
AWG10G
AWG10G
MIO
MIO: U300, U390
OUT10G
OUT10G: U500, U550
OUT10G
OUT10G: U500, U550
OUT10G
OUT10G: U500, U550
AWG10G
AWG10G: U1090
MIO
MIO: U200,U300
MIO
MIO: U200,U300
MIO
MIO: U200,U300
MIO
MIO: U200,U300
N/A
N/A
MIO, AWG10G
MIO: U300
AWG10G: U1000
MIO-AWG10G Cable
MIO, AWG10G
MIO: U300
AWG10G: U1000
MIO, AWG10G
MIO: U300
AWG10G: U1000
MIO-AWG10G Cable
MIO, AWG10G
MIO: U300
AWG10G: U1000
MIO-AWG10G Cable
N/A
N/A
MIO, AWG10G
MIO: U300
AWG10G: U1000, U400xx
MIO, AWG10G
MIO: U300
AWG10G: U1000, U400xx
MIO, AWG10G
MIO: U300
AWG10G: U1000, U400xx
MIO, AWG10G
MIO: U300
AWG10G: U1000, U400xx
N/A
AWG7000 Series Service Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

Awg7101Awg7102Awg7051Awg7052

Table of Contents