Replacing boards / replacing components
I.I.
Mount the I.I. back on the C-arm flange. The labels should face the C-arm.
Reinstall the attachment screws in the threaded holes. Turn them at least 5 revolutions.
Holding the I.I. firmly, swivel the C-arm so that the I.I. is above and the SIREPHOS is be-
low.
Carefully lower the I.I. into the service position.
I.I. Mini voltage supply
Reinstall the high voltage section of the I.I. mini voltage supply.
Reconnect the control section, however, do not tighten the screws yet.
Route the cables to the high voltage section and the control section and secure them
with cable ties or cable clamps.

VIDEOMED DC

Rotate the Compact optics into the 0° position (slowly!).
Connect an ohmmeter to the potentiometer for camera rotation, point 2 and 3.
At 0° , there should be a resistance value of 5000 ohms, + 40 ohms.
Place the VIDEOMED DC in the 0° position on the compact optics and secure it with the
clamp. Refer also to chapter 6, Replacing the VIDEOMED DC.

Miscellaneous parts

Mount the grid to the new I.I. Attach the wires for the cassette contact.
Reroute and reconnect all the cables (protective conductor, cables etc....).

Checks and adjustments

Check the I.I. electrode voltages according to the test protocol for the I.I. and adjust
them.
Check the resolution. If necessary, adjust the optical sharpness on the Compact optics.
When doing this, use the I.Q. Quick test, chapter: Resolution for evaluating the optical
sharpness.
Check the dose rate and, if necessary, adjust it.
Check that the camera optics are centered to the I.I. output and, if necessary, adjust
them. Refer also to chapter 6: Replacing the Compact optics, paragraph "Checking the
camera optics centering at the I.I. output" and "Centering the Compact optics to the I.I.
output".
Check the camera rotation and, if necessary, adjust it.
Siemens AG
Medical Solutions
SPR2-230.061.01
Rev. 13
12.04
Page 15 of 48
SIREMOBIL Iso-C
CS PS 24
6 - 15
6
6
6
6
6
System Manual