Overview Of Siremobil Iso-C Parameters; Overview Of Memoskop C / C-E / C-Sub / C & Mod / C-Sub & Mod Parameters - Siemens SIREMOBIL Iso-C Service Instructions Manual

Hide thumbs Also See for SIREMOBIL Iso-C:
Table of Contents

Advertisement

5 - 26

Overview of SIREMOBIL Iso-C Parameters

Invoke the Moreinfo.hlp help file on the service PC for a short description of the parame-
ters available on the SIREMOBIL Iso-C.
NOTE
Overview of MEMOSKOP C / C-E / C-SUB / C & MOD / C-SUB &
MOD Parameters
User Setup
The User Setup is invoked with the CTRL + U key combination.
The available functions are described in the SIREMOBIL Iso-C Operating Instructions.
Technical Setup
The technical setup is invoked with the CTRL + T key combination.
Enter the password and confirm with [RETURN]
The following functions can be invoked depending on the type of MEMOSKOP:
Function:
Test pattern generation
Video frequency
Number of monitors
Hardcopy setup
Edge 1 coef set
Edge 2 coef set
Motion detection setup
Circular mask setup
SIREMOBIL Iso-C
System Manual
If parameters have been changed and saved in the
"Adjustments, Parameter" menu with Save to file, you must exit
the service program with "Logoff" and "Quit" to save the modified
parameters to the disk.
Comments:
Not for service, invoke the necessary test images on the
service PC.
The currently programmed frame rate is shown in the
parentheses to the right. After downloading the
MEMOSKOP or after replacement, the MEMOSKOP will
start up at the default video frequency (50 Hz).
If the monitors are configured for a different video fre-
quency, you must perform "blind" configuration for the cor-
rect video frequency.
Refer to Configuring video frequency.
Allows you to program the number of monitors used (1 or 2)
You can configure the connection for hardcopy cameras.
Check whether 5000 is programmed for "4 Max Hardcopy
time" and if not, program this value.
Allows configuration of edge enhancement. Check whether
3 is programmed and if not, program this value.
Allows Configuration of edge enhancement. Check whether
8 is programmed and if not, program this value.
Configures the motion detector.
Check if at MD 1: Threshold 100 and at MD 2:Threshold
120 is programmed, if not, program these values.
Allows you to configure the diameter of the circular mask.
SPR2-230.061.01
Rev. 13
Adjustments / Programs
Page 26 of 30
12.04