Replacing boards / replacing components
Roadmap K-Factor, phase B Program "MD 2".
Roadmap K-Factor, phase C Program "MD 2".
Roadmap Time of phase A
•
Configure the following parameters in User setup for MEMOSKOP:
Operating mode Parameters Change the programmed parameters only upon customer
VCR output
Hardcopy Release
Text monitor
Split screen
Display of time
Set Date Time
Hospital Name
Storage of Dose on disk
Dose on Hardcopy output
•
Perform the IQ quick test.
NOTE
Siemens AG
Medical Solutions
Program a different value only upon customer request.
Program a different value only upon customer request.
Program "4,5".
Program a different value only upon customer request.
request.
Configure to monitor A.
For SIREMOBIL with 1 monitor: configure to monitor A.
For SIREMOBIL with 2 monitors: Configure to monitor B.
For SIREMOBIL with 1 monitor: configure to monitor A.
For SIREMOBIL with 2 monitors:
configure to both monitors.
Configure horizontally.
Configure vertically only upon customer request.
Configure Yes.
Set the current date and time.
Program the hospital name.
If area dose product meter is present, (Diamentor) con-
figure Yes. Otherwise configure No.
If area dose product meter is present, (Diamentor) con-
figure yes. Otherwise configure No.
With present System MODULARIS the parameters for the Litho-
program have to be set again according the Service instructions
LITHOSTAR MODULARIS.
SPR2-230.061.01
Rev. 13
Page 33 of 48
12.04
CS PS 24
6 - 33
SIREMOBIL Iso-C
System Manual