Adjustments / Programs
Function:
Move text positions
Date / Time format
System configuration
Keyboard
Graphics level
Display of real time clock
Display of rotated images
Product name
Diagnostics
Create External Disk (1)
Enable Bridge (1)
Delay 3D-PC/ DICOM
Bridge (1)
Interface to 3D-PC
Siemens AG
Medical Solutions
Comments:
Allows you to configure the position of superimposed text.
European: TT MM JJ / HH MM
US:
MM DD YY / HH MM
Allows you to read out the current MEMOSKOP configura-
tion.
Allows you to program the corresponding keyboard layout
Universal / Roman.
Level (brightness) of superimposed text.
(Default: 128)
Yes: Real time clock (date and time) are displayed depend-
ing on the User setup programming for MEMOSKOP.
The real time clock is active.
NO: The real time clock is switched off. The User setup pro-
gramming will not affect it.
Yes: When pressing the image rotation key, the current
image will be computerized and displayed.
NO: The rotated image will not be computerized when
pressing the image rotation key. For SIREMOBIL Iso-C
always set to YES.
Allows you to change the MEMOSKOP name, displayed
during system booting.
Enter SIREMOBIL Iso-C.
Allows you to invoke test routines for repair by the manu-
facturer of the MEMOSKOP.
The MEMOSKOP can be reset via "Load factory default",
as required. Subsequently you have to program accord-
ingly. The same applies if you replace the MEMOSKOP
MEMOSKOP CE and MEMOSKOP CE-100:
Program the value to "No".
All other types of MEMOSKOP:
If a MOD drive is implemented and/or the option
"3D Reconstruction" is installed and/ or the option DICOM
Bridge. Is installed, program the value to "Yes".
MEMOSKOP CE and MEMOSKOP CE-100:
Program the value to "No".
All other types of MEMOSKOP:
If a MOD drive is implemented and/or the option
"3D Reconstruction" is installed and/or the option DICOM
Bridge. Is installed, program the value to "Yes".
Only available if option "3D Reconstruction" is installed.
Configure 120.
Only available if option "3D Reconstruction" is installed.
Configure Yes.