Replacing boards / replacing components
Monitor zoom
Monitor split
Monitor text
Maximum image number
Disable doserate selection
in user setup
Unit serial number
Product name
Default program change-
able
Create external disk
Default LUT
Default Edge
Software backward com-
patible
•
Configure the following parameters in User setup for MEMOSKOP:
Operating mode Para-
meters
VCR output
Hardcopy Release
Text monitor
Split screen
Display of time
Set Date Time
Hospital Name
Storage of Dose on disk
Dose on Hardcopy output
•
Perform IQ quick test.
Siemens AG
Medical Solutions
Configure No.
Configure Yes.
Configure Yes.
Configure 100.
Set to "No" when equipped with an area dose product mea-
surement device. The menu "Storage of Dose on disk" and
the menu "Dose on Hardcopy output" in User Setup are visi-
ble. Otherwise set to "Yes".
Program the serial number of the SIREMOBIL.
Do not enter the leading zero.
Program "SIREMOBIL Iso-C".
Default: Yes.
Depending on country specific regulations configure Yes or
No.
Configure No.
Configure 2. Program a different LUT curve only upon cus-
tomers request.
Configure 1. Program a different Edge enhancement value
only upon customers request.
Configure No.
Change the programmed parameters only upon customer
request.
Configure to monitor A.
For SIREMOBIL with 1 monitor: configure to monitor A.
For SIREMOBIL with 2 monitors: Configure to monitor B.
For SIREMOBIL with 1 monitor: configure to monitor A.
For SIREMOBIL with 2 monitors: configure to both monitors.
Configure horizontally. Configure vertically only upon cus-
tomer request.
Configure Yes.
Set the current date and time.
Program the hospital name.
If area dose product meter is present, (Diamentor) con-
figure Yes. Otherwise configure No.
If area dose product meter is present, (Diamentor) con-
figure yes. Otherwise configure No.
SPR2-230.061.01
Rev. 13
12.04
Page 27 of 48
SIREMOBIL Iso-C
CS PS 24
6 - 27
System Manual