Memoskop C-Sub / Memoskop C-Sub & Mod - Siemens SIREMOBIL Iso-C Service Instructions Manual

Hide thumbs Also See for SIREMOBIL Iso-C:
Table of Contents

Advertisement

Replacing boards / replacing components
Dose on Hardcopy output
Perform the IQ quick test.
NOTE
MEMOSKOP C-SUB / MEMOSKOP C-SUB & MOD
NOTE
NOTE
Check the software status of the MEMOSKOP. Download the system software to the
MEMOSKOP if necessary.
Use the MEMOSKOP Service Software to import the language file.
Replace the MEMOSKOP C-SUB.
The attachment screws are accessible on the underside of the storage device.
Set the MEMOSKOP C-SUB to the correct video frequency and configure it for the num-
ber of existing monitors.
Refer to chapter 5 "Configuring the video frequency " and "MEMOSKOP - Configuration
for the number of existing monitors ."
Configure the following parameters in Technical Setup for Memoskop:
Number of Monitors
Hardcopy
Edge 1 coef set
Edge 2 coef set
Motion detection setup
Date / Time format
Keyboard
Siemens AG
Medical Solutions
If area dose product meter is present, (Diamentor)
configure yes. Otherwise configure No.
With present System MODULARIS the parameters for the Litho-
program have to be set again according the Service instructions
LITHOSTAR MODULARIS.
Check the software status of the newly delivered MEMOSKOP.
If the software status of the MEMOSKOP differs from that of the
system, the system software status must be downloaded to the
MEMOSKOP.
The MEMOSKOP software status can be read out in the 2nd line
of the User Setup.
The first number describes the MEMOSKOP type.
(2=CE; 3= CE-100/ C/ C & MOD; 4=C-SUB/ C-SUB & MOD).
Read and record the parameters for MEMOSKOP C-SUB if you
can select User setup. After replacing the MEMOSKOP C-SUB,
according to the parameters recorded.
Configure 1 for SIREMOBILwith 1 monitor.
Configure 2 for SIREMOBIL with 2 monitors.
Check whether 5000 is programmed in "4 Max Hardcopy
time". If not, program this value.
Check whether 3 is programmed, If not, program this value.
Check whether 8 is programmed, If not, program this value.
Check whether the following are programmed "MD1:
Threshold" 100 and for "MD2: Threshold" 120. If not, pro-
gram these values.
Set the country-specific format.
Set to the keyboard layout being used.
SPR2-230.061.01
Rev. 13
Page 31 of 48
12.04
CS PS 24
6 - 31
6
SIREMOBIL Iso-C
System Manual

Advertisement

Table of Contents
loading