Setting The Maximum Skin Dose Rate; Prerequisites; Preparations; Measurement - Siemens SIREMOBIL Iso-C Service Instructions Manual

Hide thumbs Also See for SIREMOBIL Iso-C:
Table of Contents

Advertisement

Adjustments / Programs

Setting the maximum skin dose rate

NOTE

Prerequisites

You must program one of the FL curves S2, HC2 or IODINE.
The FL curves S2, HC2 or IODINE should be used for the following adjustment.
The X-ray tube cover must be installed.
Use the dose ionization chamber of the dose measurement device to measure dose
rates greater than 10R/min.
If you are using an area dose product measurement device, mount the dose measure-
ment chamber above the radiation exit port of the SIREPHOS.

Preparations

Attach the dose ionization chamber
at a distance of 30 cm from the I.I.
housing (Fig. 2).
Select one of the FL curves S2 / HC2
or IODINE according to the program-
ming. (Push-key)
Briefly release fluoro.
Select kV stop mode and enter
110 kV (5 mA).
NOTE

Measurement

Switch on fluoroscopy.
Read and record the value displayed on the dose rate meter.
If the value measured exceeds 9 R/min, the maximum tube current must be reduced. If
the value measured is less than 9 R/min, the tube current does not need to be reduced.
Switch fluoroscopy off.
Siemens AG
Medical Solutions
Setting the maximum skin dose rate is necessary only for coun-
tries where DHHS regulations apply.
5
If the maximum tube current was already reduced, the mA value
displayed at 110kV should be recorded and used for the calcula-
tion. The maximum skin dose rate to be set should equal 9R/min.
SPR2-230.061.01
Rev. 13
30 cm
Chamber
Sirephos
Fig. 2
Page 9 of 30
12.04
CS PS 24
5 - 9
Input
screen
Holder
SIREMOBIL Iso-C
System Manual
5
5
5