6 - 30
Disable doserate selection
in user setup
Unit serial number
Default LUT
Default Edge
Software backward com-
patible
Product name
Default program change-
able
Create external disk
Enable Bridge
Delay 3D-PC/ DICOM
Bridge
Interface to 3D-PC
Configure the following parameters in User setup for MEMOSKOP:
Operating mode Para-
meters
VCR output
Hardcopy Release
Text monitor
Split screen
Display of time
Set Date Time
Hospital Name
Storage of Dose on disk
SIREMOBIL Iso-C
System Manual
Replacing boards / replacing components
Set to "No" when equipped with an area dose product mea-
surement device. The menu "Storage of Dose on disk" and
the menu "Dose on Hardcopy output" in User Setup are visi-
ble. Otherwise set to "Yes".
Program the serial number of the SIREMOBIL.
Do not enter the leading zero.
Configure 2. Only upon customers request program a
different LUT curve.
Configure 1. Only upon customers request program a
different Edge enhancement value.
Configure No.
Program "SIREMOBIL Iso-C".
Default: Yes.
Depending on country specific regulations configure Yes or
No.
If the option "3D Reconstruction" or the option
"DICOM Bridge" is installed or a MOD drive is present, con-
figure Yes. Otherwise configure No.
If the option "3D Reconstruction" or the option
"DICOM Bridge" is installed or a MOD drive is present, con-
figure Yes. Otherwise configure No.
Only available of if option "3D Reconstruction" is installed.
Configure 120.
Only available if option "3D Reconstruction" is installed.
Configure yes.
Change the programmed parameters only upon customer
request.
Configure to monitor A.
For SIREMOBIL with 1 monitor: configure to monitor A.
For SIREMOBIL with 2 monitors: Configure to monitor B.
For SIREMOBIL with 1 monitor: configure to monitor A.
For SIREMOBIL with 2 monitors:
Configure to both monitors.
Configure horizontally. Configure vertically only upon cus-
tomer request.
Configure Yes.
Set the current date and time.
Program the hospital name.
If area dose product meter is present, (Diamentor)
configure Yes. Otherwise configure No.
SPR2-230.061.01
Rev. 13
Page 30 of 48
12.04
CS PS 24