Disassembly
Fig. 4

X-ray image intensifier and I.I. mini voltage supply

Move the X-ray image intensifier into service position. (See chapter 1 and Fig. 4 in the
service manual.)
Disconnect the I.I. mini voltage supply (high-voltage section and control section).
(See chapter 6 in the service manual. Please observe the safety information on dis-
charging the I.I. mini voltage supply.)
Disconnect the HV cable, as well as E3, E2 and E1 cables.
Disconnect the CON1 connector.
Eject the black cathode cable of the X-ray image intensifier (CON1.8)
Remove the protective ground wire from the X-ray image intensifier and disconnect the
I.I. mini voltage supply.
Disconnect the M4 X1 camera plug from the support bracket, and remove it.
Disconnect the M30 X1 plug.
Loosen the cable ties. All image intensifier cables are now disconnected.
Remove the X-ray image intensifier from the C-arm, and store it in a dark place
(avoid exposure to direct sunlight). Avoid exposing the CCD sensor to light and dirt by
covering the camera PC board.
Remove the I.I. mini voltage supply from the C-arm.
The X-ray image intensifier section of the C-arm contains a round LEMOSA plug con-
nection. Loosen this pug connection (Fig. 5).