Replacing boards / replacing components
Maximum image number
Disable doserate selection
in user setup
Unit serial number
Product name
Default program change-
able
Software backward com-
patible
Configure the following parameters in User setup for MEMOSKOP:
Operating mode Para-
meters
VCR output
Hardcopy Release
Text monitor
Split Screen
Display of time
Set Date Time
Hospital Name
Storage of Dose on disk
Dose on Hardcopy output
Perform the IQ quick test.
NOTE
Siemens AG
Medical Solutions
n.a.
Program for "No" when equipped with an area dose product
measurement device. The menu "Storage of Dose on disk"
and the menu "Dose on Hardcopy output" in User Setup are
visible.
Otherwise set to "Yes".
Program the serial number of the SIREMOBIL.
Do not enter the leading zero.
Program "SIREMOBIL Iso-C".
Default: Yes.
Depending on country specific regulations configure Yes or
No.
Configure "No".
Change the parameters programmed for the organ program
upon customer request.
Configure to Monitor A.
For SIREMOBILwith 1 monitor: Configure to monitor A.
For SIREMOBIL with 2 monitors: Configure to monitor B.
For SIREMOBIL with 1 monitor: Configure to monitor A.
For SIREMOBIL with 2 monitors:
Configure to both monitors.
Configure horizontally. Configure vertically only upon cus-
tomer request.
Configure Yes.
Set the current date and time.
Program the hospital name.
If the area dose product meter is present, (Diamentor) con-
figure Yes. Otherwise, configure No.
If the area dose product meter is present, (Diamentor) con-
figure Yes. Otherwise configure No.
With present System MODULARIS the parameters for the Litho-
program have to be set again according the Service Instructions
LITHOSTAR MODULARIS.
SPR2-230.061.01
Rev. 13
Page 25 of 48
12.04
CS PS 24
6 - 25
SIREMOBIL Iso-C
System Manual