Replacing boards / replacing components
1
Horizontal line
Fig. 25

Angulation part

NOTE
NOTE
If the toothed wheel (1/Fig. 25) was turned several times during transport, prior to instal-
lation the perpendicular must be turned in one direction until the markings coincide again
when the angulation part is vertical.
Move the C-arm to the 0° position.
Disconnect the cable of the actual value potentiometer.
Loosen the fastening screws of the angulation part.
Replace the angulation part. In doing so, ensure that, when the C-arm is perpendicular,
the markings (black lines) on the two toothed wheels coincide.
Reconnect the cable of the actual value potentiometer.
Perform the mechanical settings according to the Startup Instructions, Chapter 6.
Siemens AG
Medical Solutions
Marker
Perpendicular line
Only applies to SIREMOBIL Iso-C 3D.
When the angulation part is vertical, the markings on the two
toothed belts must coincide (see Fig. 25).
SPR2-230.061.01
Rev. 13
12.04
Page 47 of 48
SIREMOBIL Iso-C
CS PS 24
6 - 47
6
System Manual