Instructions For Use - Otto Bock 4R69-AL Instructions For Use Manual

Hide thumbs Also See for 4R69-AL:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 7
► Take any necessary measures (e.g. repair, replacement, inspec­
tion by the manufacturer's customer service, etc.).
NOTICE
Mechanical overload
Impaired functionality due to mechanical damage
► Check the product for damage prior to each use.
► Do not use the product if its functionality has been impaired.
► Take any necessary measures (e.g. repair, replacement, inspec­
tion by the manufacturer's customer service, etc.).
Signs of changes in or loss of functionality during use
Among other factors, changes in functionality can be indicated by an
altered gait pattern, a change in the positioning of the prosthetic com­
ponents relative to each other and by the development of noises.
4 Scope of Delivery
The following single components and accessories are included in the
scope of delivery in the specified quantities and are available for
reordering as single components ( ), single components with minim­
um order quantities (
) or single-component packs ( ):
Fig.
Item
Quant­
Designation
ity
1 (   )

Instructions for use

1
Tube clamp adapter
1 ( 
 ) Cap screw
4 ( 
 ) Set screw
5 Preparation for Use
CAUTION
Incorrect alignment or assembly
Injuries due to fall resulting from damaged prosthesis components
8 | Ottobock
► Observe the alignment and assembly instructions.
CAUTION
Improper assembly of the screw connections
Fall due to breakage or loosening of the screw connections
► Clean the threads before every installation.
► Apply the specified installation torque values.
► Observe the instructions for securing the screw connections and
the use of the correct length.
5.1 Mounting the adapter
> Recommended tools and materials:
710D4 Torque Wrench, degreasing cleaner (e.g. 634A3 Acetone)
1) Loosen the tube clamp cap screw by 2 turns.
2) For final assembly: clean the insertion range of the tube clamp
adapter and tube adapter with degreasing cleaner.
3) Fully insert the tube adapter into the tube clamp adapter up to the
stop. Align the clamping slot:
Tube clamp adapter: anterior
Tube clamp adapter, sliding: anterior or medial
4) Tighten the tube clamp cap screw (installation torque: M4: 5 Nm,
Reference num­
M5: 10 Nm, M6: 13 Nm).
ber
5.2 Installation in the modular prosthesis
> Required tools and materials:
710D4 torque wrench, 636K13 Loctite®
501Z2=M6x25
► Install the other prosthetic components as described in the
respective instructions for use.
506G3=M8x14
6 Cleaning
1) Clean the product with a damp, soft cloth.
2) Dry the product with a soft cloth.
3) Allow to air dry in order to remove residual moisture.

Advertisement