Technical specifications
13.12 Ventilator
Ventilator
Type
Patient range
Ventilation modes
Tidal volume range
(volume controlled modes)
Tidal volume setting range
Minute volume setting range
Inspiratory pressure
(Pressure controlled modes)
Inspiratory pressure
setting range
Compressible volume
compensation
Max. inspiratory flow
Breathing frequency
I:E (VC, PC)
PEEP
Trigger
Inspiratory pause (VC)
276
|
Pneumatic powered Servo controlled
Neonatal to Adult
Manual/Bag
AFGO (option)
Volume Control (VC)
Pressure Control (PC)
Pressure Support (PS, option)
20 – 2000 ml, ±10% or 10 ml, whichever is greater
Infant range:
Adult range:
Infant range:
Adult range:
0 - 120 cmH
Infant range:
Adult range:
Yes
200 l/min (3.3 l/s)
4 – 100 ±1 breaths/minute
1:10 - 4:1
0 – 50 cmH
2
Flow / Pressure
0 to 30% or 0 - 1.5 s
20 - 350 ml, resolution 1 ml
100 - 2000 ml, resolution 10 ml
0.3 - 20 l/min
0.5 - 60 l/min
O
± 15% or ± 2 cmH
2
greater
0 - 80 cmH
0 - 120 cmH
O
O, whichever is
2
O, resolution 1 cmH
2
2
O, resolution 1 cmH
2
FLOW-i 4.2, User's Manual
O
O
2