Getinge Servo-u Cleaning Training
Hide thumbs Also See for Servo-u:

Advertisement

Quick Links

MX-6989-02
MCV00052945 REVB
Servo-u/n
Cleaning

Advertisement

loading
Need help?

Need help?

Do you have a question about the Servo-u and is the answer not in the manual?

Questions and answers

Summary of Contents for Getinge Servo-u

  • Page 1 Servo-u/n Cleaning MX-6989-02 MCV00052945 REVB...
  • Page 2 Table of contents General Cleaning Expiratory cassette MX-6989-02 MCV00052945 REVB...
  • Page 3 General Different levels of cleaning Clean To free from dirt or pollution Clean Disinfect To reduce the number of harmful microorganisms Sterilize To free from living microorganisms Sterilize Disinfect MX-6989-02 MCV00052945 REVB...
  • Page 4 General Infected parts Cautions: All personnel should be aware of the risk of parts being infected when disassembling and cleaning the ventilator system. MX-6989-02 MCV00052945 REVB...
  • Page 5 General Cleaning intervals The cleaning methods vary between different hospitals. MX-6989-02 MCV00052945 REVB...
  • Page 6 General Servo-u/n cleaning instructions User’s manual SERVO-U ® / SERVO-n ® Cleaning and maintenance MX-6989-02 MCV00052945 REVB...
  • Page 7 Table of contents General Cleaning Expiratory cassette MX-6989-02 MCV00052945 REVB...
  • Page 8 Cleaning Diagrams/Wipe off/discard Cleaning when Disinfection when no Steam autoclave bacterial filter is used bacterial filter is used Dismantle Dismantle Dismantle Wipe off/ Wipe off/ Wipe off/ discard discard discard Clean/Rinse Rinse Assemble Washer- Disinfectant disinfector Autoclave Rinse Assemble Assemble MX-6989-02 MCV00052945 REVB...
  • Page 9: Cleaning Diagram

    Cleaning diagram Wipe off/discard Wipe the ventilator system and all removable parts with cleaning wipes or a soft lint-free cloth moistened in soap & water or detergent-based disinfectant •  3M Respirator Cleaning Wipes •  Clorox Bleach Germicidal Wipes (may cause discoloration) • ...
  • Page 10 Table of contents General Cleaning Expiratory cassette MX-6989-02 MCV00052945 REVB...
  • Page 11 Expiratory cassette cleaning When bacterial filter is used The need for cleaning of the expiratory cassette is heavily reduced by using an Dismantle external bacterial filter. Disconnect and change the expiratory Wipe off / discard filter: Assemble •  Follow manufacturer’s instructions on how long the filter can be used e.g.
  • Page 12 Expiratory cassette cleaning Disinfection When no bacterial filter is used, disinfection is necessary. Dismantle Wipe off/ discard Rinse Washer- Disinfectant disinfector Rinse Assemble MX-6989-02 MCV00052945 REVB...
  • Page 13 Expiratory cassette cleaning Rinse Immediate rinsing in water If rinsing before disinfection Dismantle (<35°C/95°F) is not included in the can remove particles washer disinfector Wipe off/ and reduces the risk of program: discard cross-contamination •  Rinse the parts between patients. thoroughly in water Rinse (<35°C/95°F).
  • Page 14 Expiratory cassette cleaning Pre-disinfectant •  Alcohol (ethyl or isopropyl alcohol 70%) •  Aniosyme DD1 •  Gigazyme Plus •  Salvanios pH7 (new instead of Hexanios) Important: Follow the disinfectant manufacturer’s recommendations and instructions. MX-6989-02 MCV00052945 REVB...
  • Page 15 Expiratory cassette cleaning Rinse •  Rinse the parts Dismantle thoroughly in water to remove all traces of disinfectant Wipe off/ discard •  Let the water flow Rinse through the parts •  Carefully shake Washer- Disinfectant and tilt the cassette, turn disinfector upside down Rinse...
  • Page 16 Expiratory cassette cleaning Washer-disinfector The expiratory cassette will last at least 200 cycles. Dismantle Wipe off/ discard Rinse Washer- Disinfectant disinfector Rinse Assemble MX-6989-02 MCV00052945 REVB...
  • Page 17 •  Wash the parts with water only in a Dismantle washer-disinfector at a temperature of 85-95°C (185-203°F) Wipe off/ discard •  Cleaning agent that can be used: Rinse -  neodisher ® MediClean forte -  Getinge MIS Detergent Washer- Disinfectant disinfector Rinse Assemble MX-6989-02 MCV00052945 REVB...
  • Page 18 Expiratory cassette cleaning Disinfectant • Alcohol (ethyl or isopropyl alcohol 70%) • Anioxyde 1000 • Cidex OPA Important: Follow the disinfectant manufacturer’s recommendations and instructions. MX-6989-02 MCV00052945 REVB...
  • Page 19 Expiratory cassette cleaning Drying Drying cabinet Room air Dismantle Wipe off/ discard Rinse Washer- Disinfectant disinfector Rinse Important: Never dry the cassette by applying high pressure air. Internal tubing may be damaged. Assemble MX-6989-02 MCV00052945 REVB...
  • Page 20 Expiratory cassette cleaning Steam autoclave •  Steam autoclaving will reduce the Dismantle lifetime of the expiratory cassette and is not recommend as a cleaning method Wipe off/ discard •  Before placing the expiratory cassette in a steam autoclave, make sure that no Clean/Rinse water remains inside the cassette Autoclave...
  • Page 21 Expiratory cassette cleaning Steam autoclave •  Instrument parts should be steam Dismantle autoclaved in a validated process, at a temperature of 134°C (273°F) for 3 Wipe off/ minutes. discard •  Rubber parts should be steam Clean/Rinse autoclaved in a validated process, at a temperature of 121°C (250°F) for 15-20 minutes.
  • Page 22 Cleaning the expiratory cassette Steam autoclave The expiratory cassette will last at least: •  100 autoclaving cycles with 3 minutes sterilization time at 134ºC (273ºF) •  30 autoclaving cycles with 18 minutes sterilization time at 134ºC (273ºF) MX-6989-02 MCV00052945 REVB...
  • Page 23 Cleaning the expiratory cassette Pre-use check A pre-use check must be performed after all types of cleaning. If the pre-use check fails: •  Carefully shake/tilt the cassette (5-7 times) •  Run the cassette in a ventilator system with a test lung for 10 minutes • ...
  • Page 24 Getinge is a leading global provider of innovative solutions for operating rooms, intensive-care units, hospital wards, sterilization departments, elderly care and for life science companies and institutions. With a genuine passion for life we build quality and safety into every system. Our unique value proposition mirrors the continuum of care, enhancing efficiency throughout the clinical pathway. Based on our first-hand experience and close partnerships, we are able to exceed expectations from customers –...