User Calibration - Feasa 3D User Manual

Display analyser
Table of Contents

Advertisement

FEASA DISPLAY ANALYSER
FUNCTIONAL VERSION

User Calibration

The Display Analyser must be calibrated with a known measured Display to
obtain the best results from the Analyser.
It is recommended to measure the Luminance of the Display with the Feasa
LED Spectrometer fitted with the Luminance Head.
The Display Luminance should be measured at the same points that will be
measured with the Display Analyser because the Luminance of the Display can
vary across the surface.
It is also advised to let the Display warm up until the Luminance stabilizes.
This can be monitored with the Spectrometer and Luminance head.
Calibration of the Display is carried out using the Feasa UserCal program which
is supplied on the accompanying CD.
Up to 10 calibration sets can be stored in the Analyser in addition to the Factory
default setting. This allows up to 10 different Displays to be tested.
The following commands allow the User to select different Calibration sets.