Setyoffset##0.Yyy - Set The Y Chromaticity Offset - Feasa 3D User Manual

Display analyser
Table of Contents

Advertisement

setYOFFSET##0.yyy
Transmit
Setyoffset##±0.yyy
Where:
##
represents the Fiber Number and is a number in the range 01 – 20.
0.yyy
represents the y Chromaticity offset value ±(0.000 – 0.300).
Description
This command is used to set an offset to the displayed y Chromaticity value.
The limit of the offset is ±0.300 which means values must be in the range
-0.300 to 0.000 to 0.300. This command is useful when the user wishes to set
the y Chromaticity to be the same as that specified by the LED Manufacturer.
The default value of the offset is 0.000.
memory and will remain at the programmed setting until changed by a new
Setyoffset command.
Example:
The PC transmits setyoffset01-0.050 to the LED Analyser to instruct it to set
the y offset on Fiber 01 to -0.050.
setyoffset01-0.050
OK
© Copyright Feasa Enterprises Ltd
FEASA DISPLAY ANALYSER
FUNCTIONAL VERSION
- Set the y Chromaticity Offset
Set Data Mode
Receive
OK
The offset is stored in non-volatile
Back to Index
38