FEASA DISPLAY ANALYSER
FUNCTIONAL VERSION
Get Data Mode
getAbsInt##
Get the Absolute Intensity Value of the Display under test.
Transmit
Receive
getABSINT##
#.####e±##
Where:
##
represents the Channel No in the range 01 – 10.
#.####e±## represents the data in Scientific format.
Description:
This command will instruct the Display Analyser to return the Absolute Intensity
value for the Display positioned in front of Channel##.
Absolute Intensity values must first be measured with the Feasa LED
Spectrometer. The Display Analyser must then be calibrated with these values
using the programs supplied with the Spectrometer.
This command should be preceded by a Capture command to ensure valid
data is stored in the memory of the Display Analyser.
The data for each
Channel can be read out one-at-the-time, in any order.
Note: it is important
for the PC to use 01, 02, 03, etc to indicate Fibers 1, 2 and 3, etc.
Example:
The PC transmits getabsint01 to the Display Analyser to instruct it to send the
stored absolute Intensity data for the Display positioned in front of Channel No
1.
The Display Analyser will return a string in the format #.####e±## to
the PC.
getabsint01
2.5000e-02
Back to Index