Get Data Commands

he get data commands are used to read out the Color, Saturation and
T
Intensity data stored by the capture commands.
The data from the last capture command remains in memory until a new
capture command is issued or the power is removed from the Analyser.
Commands are transmitted and received using ASCII characters and are NOT
case-sensitive.
character.

Under Range Condition

An under range condition will occur when insufficient light from the LED
reaches the sensor for the range selected. This will be indicated by 000 000
000 00000 for RGBI and 0.0000 0.0000 for xy and uv.
If this condition occurs select the next lower range and test again.

Over Range Condition

An over range condition will occur when too much light from the LED reaches
the sensor for the range selected by the switch. This will be indicated by 255
255 255 99999 for RGBI and 0.0000 0.0000 for xy and uv.
If this condition occurs select the next higher range and test again.
© Copyright Feasa Enterprises Ltd
FEASA DISPLAY ANALYSER
FUNCTIONAL VERSION
All commands must be terminated with a <CR> or <LF>
Capture Mode
Back to Index